Process Loop Powered Indicators

  • 3-1/2 Digit Display with Dummy Zero
  • Powered Entirely by Signal Loop
  • 4 to 20 or 10 to 50 mA Ranges
  • Scaleable Zero and Offset for Engineering Units
  • Optional CENELEC Certification

    The 508B/558B Series two-wire current-loop indicators accept 1-5, 4-20 or 10-50 mA signals and digitally display the process variable in percent or in engineering units such as pressure, flow, temperature, and level. No separate power supply is required, as these units take power directly from the current loop with a voltage drop of less then 2.5 volts. The units have full zero and span scaling capability for any desired engineering unit display, six coarse zero ranges, three coarse span ranges, three with jumper-select decimal location, and a dummy zero.
    The 558B has a panel mount design while the 508B Series are more ruggedly designed, with the electronics securely connected to the top of the case. They are designed to withstand high vibration and shock with cases of die-cast metal that are waterproof to 35kPa (5 psi). The 508B-1 has barrier strip terminals, and the 508B-2 has a 1/2" EMT male conduit mounting.

Model 508B-1 Shown

558B
Model 558B Shown

SPECIFICATIONS
Current: 1-5 mA, 4-20 mA and 10-50 mA
Protection: 200 mA max., forward and 100 mA max. reverse
Voltage Drop: 2.5 Volts
Zero Adjust: -2000 to +2000 counts
Span Adjust: 0 to 2000 counts
Normal Mode Rejection (NMR): 46 dB, 50-60 Hz
Accuracy at 25°C: max. error: ±0.1% of span ±1 count; Zero Tempco: ±0.1 count/°C typ., ±0.3 count/°C max Span Tempco: 0.005% of span/°C typ., 0.15% of span/°C max.
Integration Period: 100 ms
Read Rate: 2.5 s
Display Type 508B: 7 segment LCD, 8.9 mm (0.35 in) high; jumper select decimal point; 558B: 7 segment LCD, 12.7 mm (0.50 in.) high
Overrange: three least-significant active digits blank
Environmental Operating Temperature:
508B:
-40 to +85°C; 558B: optional -40 to 85°C; 0 to 55°C
Mechanical (558B)
Weight: 128 g
Case Material: ABS polycarbonate
Case Size: bezel: (W x H x T): 96 x 48 x 8 mm; depth behind bezel with cover: 100 mm; panel cutout (W x H): 92 x 45 mm
Electrical Connections: plug-in socket with screw clamps for wire sizes 16 to 26 AWG
Mechanical (508B)
Weight: 400 g (14 oz.)
Diameter: 74 mm (2.9 in)
Height (incl. barrier): 48 mm (1.9 in)
Electrical Connections: 508B-1: 3-terminal barrier; 508B-3: 1/2" NPT male conduit fitting with two 0.3 meter long #18 stranded wires

To Order (Specify Model Number) Prices Shown in Pounds
Model Description Price
558B Current Loop Indicator £116
558B-ET Current Loop Indicator with extended ambient range (-40 to 85 ° C) 143
508B-1 Flat surface or relay track mounting 252
508B-2 EMT conduit mounting 252

Options

Order Suffix Description Add'l Price
-BL Blank lens without NEWPORT logo £12
-FS Custom configuration and calibration 25
-IS1 Intrinsic Safety certification, with CENELEC EEx ia certification label 22

Mounting Options, 508B-1 only

Order Suffix Description Add'l Price
-MAT1 Adaptor plate for surface mount or 69.9 mm relay track £ 6
-MDT1 Rail clamp for DIN relay track 10

Accessories

Model Description Price
MXS1 Spring retainers for 508B-1 in explosion proof housing EPW3 £ 8
MXS2 Spring retainers for 508B-1 in explosion proof housing EPW2 8
EPW3 NEMA 4 and NEMA 7 rated environmental enclosure, single height; includes MXS1 hardware 70
EPW2 NEMA 4 and NEMA 7 rated environmental enclosure, double height for 508B-1 and 500 series transmitter; includes MXS1 and MXS2 hardware 85
Comes with complete operator's manual.
Ordering Example: 508B-1 current loop indicator with flat surface or relay track mounting. £252.

 